1.

Introduction:

Preanalytical errors, along the process from the beginning of test requests to the admissions of the specimens to the laboratory, cause the rejection of samples. The aim of this study was to better explain the reasons of rejected samples, regarding to their rates in certain test groups in our laboratory.

Materials and methods:

This preliminary study was designed on the rejected samples in one-year period, based on the rates and types of inappropriateness. Test requests and blood samples of clinical chemistry, immunoassay, hematology, glycated hemoglobin, coagulation and erythrocyte sedimentation rate test units were evaluated. Types of inappropriateness were evaluated as follows: improperly labelled samples, hemolysed, clotted specimen, insufficient volume of specimen and total request errors.

Results:

A total of 5,183,582 test requests from 1,035,743 blood collection tubes were considered. The total rejection rate was 0.65 %. The rejection rate of coagulation group was significantly higher (2.28%) than the other test groups (P < 0.001) including insufficient volume of specimen error rate as 1.38%. Rejection rates of hemolysis, clotted specimen and insufficient volume of sample error were found to be 8%, 24% and 34%, respectively. Total request errors, particularly, for unintelligible requests were 32% of the total for inpatients.

Conclusions:

The errors were especially attributable to unintelligible requests of inappropriate test requests, improperly labelled samples for inpatients and blood drawing errors especially due to insufficient volume of specimens in a coagulation test group. Further studies should be performed after corrective and preventive actions to detect a possible decrease in rejecting samples.  相似文献   

2.

Background:

Presently the necessity of fasting time for coagulation tests is not standardized. Our hypothesis is that this can harm patient safety. This study is aimed at evaluating whether a light meal (i.e. breakfast) can jeopardize laboratory coagulation tests.

Materials and methods:

A blood sample was firstly collected from 17 fasting volunteers (12 h). Immediately after blood collection, the volunteers consumed a light meal. Then samples were collected at 1, 2 and 4 h after the meal. Coagulation tests included: activated partial thromboplastin time (APTT), prothrombin time (PT), fibrinogen (Fbg), antithrombin III (AT), protein C (PC) and protein S (PS). Differences between samples were assessed by Wilcoxon ranked-pairs test. The level of statistical significance was set at P < 0.05. Mean % differences were determined and differences between and baseline and 1, 2 and 4h samples were compared with reference change value (RCV).

Results:

A significantly higher % activity of AT was observed at 1 h and 4 h after meal vs. baseline specimen [113 (104–117) and 111 (107–120) vs. 109 (102–118), respectively; P = 0.029 and P = 0.016]. APTT at 2 h was found significantly lower than baseline samples [32.0 (29.9–34.8) vs. 34.1 (32.2–35.2), respectively; P = 0.041]. The results of both Fbg and PS tests were not influenced by a light meal. Furthermore, no coagulation tests had significant variation after comparison with RCV.

Conclusion:

A light meal does not influence the laboratory coagulation tests we assessed, but we suggest that the laboratory quality managers standardize the fasting time for all blood tests at 12 hours, to completely metabolize the lipids intake.  相似文献   

3.

Introduction

The emergency laboratory in Hacettepe University Hospitals receives specimens from emergency departments (EDs), inpatient services and intensive care units (ICUs). The samples are accepted according to the rejection criteria of the laboratory. In this study, we aimed to evaluate the sample rejection ratios according to the types of pre-preanalytical errors and collection areas.

Materials and methods

The samples sent to the emergency laboratory were recorded during 12 months between January to December, 2013 in which 453,171 samples were received and 27,067 specimens were rejected.

Results

Rejection ratios was 2.5% for biochemistry tests, 3.2% for complete blood count (CBC), 9.8% for blood gases, 9.2% for urine analysis, 13.3% for coagulation tests, 12.8% for therapeutic drug monitoring, 3.5% for cardiac markers and 12% for hormone tests. The most frequent rejection reasons were fibrin clots (28%) and inadequate volume (9%) for biochemical tests. Clotted samples (35%) and inadequate volume (13%) were the major causes for coagulation tests, blood gas analyses and CBC. The ratio of rejected specimens was higher in the EDs (40%) compared to ICUs (30%) and inpatient services (28%). The highest rejection ratio was observed in neurology ICU (14%) among the ICUs and internal medicine inpatient service (10%) within inpatient clinics.

Conclusions

We detected an overall specimen rejection rate of 6% in emergency laboratory. By documentation of rejected samples and periodic training of healthcare personnel, we expect to decrease sample rejection ratios below 2%, improve total quality management of the emergency laboratory and promote patient safety.Key words: clinical laboratory services, total quality management, patient safety, preanalytical phase, preanalytical error  相似文献   

4.

Introduction:

There are a number of pre-analytical and analytical factors, which cause false results in the complete blood count. The present case identifies cold agglutinins as the cause for the mismatch between hematocrit and hemoglobin values.

Materials and methods:

70-year old female patient had a history of cerebrovascular diseases and rheumatoid arthritis. During routine laboratory examination, the patient had normal leukocyte and platelet counts; however, the hemoglobin (Hb: 105 g/L) and hematocrit (HCT: 0.214 L/L) results were discordant. Hemolysis, lipemia and cold agglutinin were evaluated as possible reasons for the mismatch between hematocrit and hemoglobin values.

Results:

First blood sample was slightly hemolysed. Redrawn sample without hemolysis or lipemia was analyzed but the mismatch became even more distinct (Hb: 104 g/L and HCT: 0.08 L/L). In this sample, the titration of the cold agglutinin was determined and found to be positive at 1:64 dilution ratios. After an incubation of the sample at 37°C for 2 hours, reversibility of agglutination was observed.

Conclusion:

We conclude that cold agglutinins may interfere with the analysis of erythrocyte and erythrocyte-related parameters (HCT, MCV, MCH and MCHC); however, Hb, leukocyte and platelet counts are not affected.  相似文献   

5.

Background

Platelet satellitism is a phenomenon of unknown etiology of aggregating platelets around polymorphonuclear neutrophils and other blood cells which causes pseudothrombocytopenia, visible by microscopic examination of blood smears. It has been observed so far in about a hundred cases in the world.

Case subject and methods

Our case involves a 73-year-old female patient with a urinary infection. Biochemical serum analysis (CRP, glucose, AST, ALT, ALP, GGT, bilirubin, sodium, potassium, chloride, urea, creatinine) and blood cell count were performed with standard methods on autoanalyzers. Serum protein fractions were examined by electrophoresis and urinalysis with standard methods on autoanalyzer together with microscopic examination of urine sediment. Erythrocyte sedimentation rate, blood culture and urine culture tests were performed with standard methods.

Results

Due to typical pathological values for bacterial urinary infection, the patient was admitted to the hospital. Blood smear examination revealed phenomenon, which has persisted for three weeks after the disease has been cured. Blood smears with EDTA as an anticoagulant had platelet satellitism whereas the phenomenon was not observed in tubes with different anticoagulants (Na, Li-heparin) and capillary blood.

Discussion

We hypothesize that satellitism was induced by some immunological mechanism through formation of antibodies which have mediated platelets binding to neutrophil membranes and vice versa. Unfortunately we were unable to determine the putative trigger for this phenomenon. To our knowledge this is the second case of platelet satellitism ever described in Croatia.Key words: blood platelets, thrombocytopenia, EDTA, urinary infection  相似文献   

6.

Introduction

The aim of the present study was to validate prothrombin time (PT) international normalized ratio (INR) results obtained using Steelex test reagents and a Steelex coagulometer (Steelex Scientific Instrument Company, Beijing, China), in comparison with use of a well-established standard test employing Pacific Hemostasis reagents (Fisher Diagnostics, Middletown, VA, USA) and Teco Coatron A4 coagulometer (Teco Medical Instruments GmbH, Neufahrn, Germany).

Materials and methods:

Between- and within-day coefficients of variation (CVs) of both assays were calculated using control samples provided by the test manufacturers. Samples from 90 subjects were collected and INR values were determined in a double-blind parallel manner employing both systems.

Results:

The within-day coefficients of variation (CVs) in INR estimates ranged from 2.6% (INR = 1.12) to 3.1% (INR = 2.51) for the Steelex system and from 2.1% (INR = 1.09) to 1.8% (INR = 2.8) for the Pacific test; the between-day values ran from 3.4% (INR = 1.16) to 7.9% (INR = 2.64) and from 3.3% (INR = 1.1) to 2.3% (INR = 2.7), respectively. Passing-Bablok fit of the of the Steelex and Pacific methods yielded the equation: Steelex INR = 0.85 (0.79–0.91) × Pacific INR + 0.12 (−0.02–0.21), whereas the CUSUM linearity P value was < 0.01. The mean bias as determined by the Bland-Altman test was −0.156 (−0.912–0.600).

Conclusion:

The results obtained using Steelex reagents and the M600H coagulometer are not equivalent to those obtained using Pacific Hemostasis reagents and a Teco Coatron A4 coagulometer, at least in the therapeutic range.  相似文献   

7.

Introduction

Preanalytical control and monitoring continue to be an important issue for clinical laboratory professionals. The aim of the study was to evaluate a monitoring system of preanalytical errors regarding not suitable samples for analysis, based on different indicators; to compare such indicators in different phlebotomy centres; and finally to evaluate a single synthetic preanalytical indicator that may be included in the balanced scorecard management system (BSC).

Materials and methods

We collected individual and global preanalytical errors in haematology, coagulation, chemistry, and urine samples analysis. We also analyzed a synthetic indicator that represents the sum of all types of preanalytical errors, expressed in a sigma level. We studied the evolution of those indicators over time and compared indicator results by way of the comparison of proportions and Chi-square.

Results

There was a decrease in the number of errors along the years (P < 0.001). This pattern was confirmed in primary care patients, inpatients and outpatients. In blood samples, fewer errors occurred in outpatients, followed by inpatients.

Conclusion

We present a practical and effective methodology to monitor unsuitable sample preanalytical errors. The synthetic indicator results summarize overall preanalytical sample errors, and can be used as part of BSC management system.Key words: Preanalytical phase, errors in laboratory medicine, balanced scorecard, patient safety  相似文献   

8.

Background:

A number of preanalytical activities strongly influence sample quality, especially those related to sample collection. Since blood drawing through intravenous catheters is reported as a potential source of erythrocyte injury, we performed a critical review and meta-analysis about the risk of catheter-related hemolysis.

Materials and methods:

We performed a systematic search on PubMed, Web of Science and Scopus to estimate the risk of spurious hemolysis in blood samples collected from intravenous catheters. A meta-analysis with calculation of Odds ratio (OR) and Relative risk (RR) along with 95% Confidence interval (95% CI) was carried out using random effect mode.

Results:

Fifteen articles including 17 studies were finally selected. The total number of patients was 14,796 in 13 studies assessing catheter and evacuated tubes versus straight needle and evacuated tubes, and 1251 in 4 studies assessing catheter and evacuated tubes versus catheter and manual aspiration. A significant risk of hemolysis was found in studies assessing catheter and evacuated tubes versus straight needle and evacuated tubes (random effect OR 3.4; 95% CI = 2.9–3.9 and random effect RR 1.07; 95% CI = 1.06–1.08), as well as in studies assessing catheter and evacuated tubes versus catheter and manual aspiration of blood (OR 3.7; 95% CI = 2.7–5.1 and RR 1.32; 95% CI = 1.24–1.40).

Conclusions:

Sample collection through intravenous catheters is associated with significant higher risk of spurious hemolysis as compared with standard blood drawn by straight needle, and this risk is further amplified when intravenous catheter are associated with primary evacuated blood tubes as compared with manual aspiration.  相似文献   

9.

Introduction:

The aim of this study was to investigate whether serum levels of interleukin-1β (IL–1β) has any possible correlation on inflammatory parameters such as C-reactive protein (CRP), erythrocyte sedimentation rate (ESR) and fibrinogen concentration in patients with familial Mediterranean fever (FMF) patients during attack-free period.

Materials and methods:

The serum levels of IL-1β, as an indicator of cytokines status, and the acute phase response proteins, CRP, ESR and fibrinogen levels were evaluated in 35 attack-free patients with FMF and 25 healthy volunteers.

Results:

Serum IL-1β levels were significantly higher in patients with FMF than control subjects (P = 0.018). There was no statistically significant difference in the serum levels of ESR, CRP and fibrinogen between two groups (P = 0.181, P = 0.816, P = 0.686, respectively). There was a significant correlation between IL-1β and CRP (r = 0.513, P = 0.002) values of FMF group.

Conclusions:

In conclusion, our results confirm the presence of increased IL-1β levels in FMF patients during attack-free period. Serum IL-1β values seems to correlate with CRP levels. The elevation of IL-1β levels may be important in monitoring subclinical inflammation of attack free period in FMF patients.  相似文献   

10.

Introduction:

Proper preparation of the individual is a key prerequisite for ensuring the quality of laboratory testing. Our hypothesis was that many outpatients are not sufficiently familiar with the correct way of preparing for the laboratory tests, for which the individual needs to be at fasting. This study aimed to investigate: i) whether patients are aware of how they need to prepare properly for laboratory tests; ii) the way in which users are informed about how to prepare for laboratory testing; and iii) whether users arrive to the laboratory for phlebotomy properly prepared.

Materials and methods:

An anonymous questionnaire was conducted on 150 outpatients older than 18 years, during February 2013. The response rate was 11%. All patients were interviewed by the laboratory staff. Patients were informed about detail of the questionnaire and agreed to participate in the survey.

Results:

Out of the total number subjects, 39% were fully aware of the proper definition of the fasting, whereas even 46% subjects replied that the last meal has to be taken the day before and the exact time that must pass after the last meal to blood sampling is not important. Furthermore, 52% subjects did not receive any information about how they need to prepare themselves properly for blood testing. Only 60% of them came properly prepared for the laboratory blood testing.

Conclusions:

Substantial proportion of patients do not come properly prepared for laboratory testing. We conclude that patients are not well informed about the fasting requirements for laboratory blood testing. Moreover, requesting physician is the preferred source of information from which patients learn how to prepare themselves for phlebotomy.  相似文献   

11.

Introduction:

Blood collection through intravenous lines frequently causes spurious hemolysis. Due to specific structure, the tube holder Holdex® (Greiner Bio-One GmbH, Kremsmuenster, Austria) is supposed to prevent erythrocyte injury in samples collected from catheters, so that we planned a specific study to support this hypothesis.

Materials and methods:

Blood was collected from emergency department (ED) patients with 20-gauge catheter. In patients with odd order numbers, first and second tubes were collected with conventional holder (BD Vacutainer One Use Holder, Becton Dickinson, Milan, Italy) and the third with Holdex, whereas in even patients first and second tubes were drawn with Holdex and the third using BD Vacutainer One Use Holder. The first tube was discarded, whereas the second and third were centrifuged and serum was tested for potassium, lactate dehydrogenase (LD) and hemolysis index.

Results:

The final study population consisted in 60 ED patients. Concentrations of potassium (4.25 vs. 4.16 mmol/L; P = 0.031), LD (498 vs. 459 U/L; P = 0.039) and cell-free hemoglobin (0.42 vs. 0.22 g/L; P = 0.042) were higher in samples collected with BD Vacutainer One Use Holder than with Holdex. The mean bias of cell-free hemoglobin was −0.4 g/L in samples collected with Holdex. Although the frequency of samples with cell-free hemoglobin > 0.5 g/L was identical (17/60 vs. 17/60; P = 1.00), the frequency of those with concentrations >3.0 g/L was higher using BD Vacutainer One Use Holder than Holdex (4/60 vs. 0/60; P = 0.042).

Conclusions:

The use of Holdex for drawing blood from intravenous lines may be effective for reducing gross hemolysis.  相似文献   

12.

Introduction

Measurements of blood ethanol concentrations must be accurate and reliable. The most important factors affecting blood ethanol stability are temperature and storage time. In this study, we aimed to compare ethanol stability in plasma samples at -20 °C for the different storage periods.

Materials and methods

Blood samples were collected from intoxicated drivers (N = 80) and initial plasma ethanol concentrations were measured immediately. Plasma samples were then stored at -20 °C and re-assessed after 2, 3, 4, or 5 months of storage. Differences between the initial and stored ethanol concentrations in each group (N = 20) were analyzed using Wilcoxon matched-pairs test. The deviation from the initial concentration was calculated and compared with Clinical Laboratory Improvement Amendments (CLIA’88) Proficiency Testing Limits. Relationships between the initial concentrations and deviations from initial concentrations were analyzed by Spearman’s correlation analysis. For all statistical tests, differences with P values of less than 0.05 were considered statistically significant.

Results

Statistically significant differences were observed between the initial and poststorage ethanol concentrations in the overall sample group (P < 0.001). However, for the individual storage duration groups, analytically significant decreases were observed only for samples stored for 5 months, deviations from the initial concentrations exceeded the allowable total error (TEa). Ethanol decreases in the other groups did not exceed the TEa.

Conclusion

According to our results, plasma ethanol samples can be kept at -20 °C for up to 3-4 months until re-analysis. However, each laboratory should also establish its own work-flow rules and criterion for reliable ethanol measurement in forensic cases.Key words: preanalytical phase, ethanol, stability, storage temperature  相似文献   

13.

Introduction:

The activities involving phlebotomy, a critical task for obtaining diagnostic blood samples, are poorly studied as regards the major sources of errors and the procedures related to laboratory quality control. The aim of this study was to verify the compliance with CLSI documents of clinical laboratories from South America and to assess whether teaching phlebotomists to follow the exact procedure for blood collection by venipuncture from CLSI/NCCLS H03-A6 - Procedures for the Collection of Diagnostic Blood Specimens by Venipuncture might improve the quality of the process.

Materials and methods:

A survey was sent by mail to 3674 laboratories from South America to verify the use of CLSI documents. Thirty skilled phlebotomists were trained with the CLSI H03-A6 document to perform venipuncture procedures for a period of 20 consecutive working days. The overall performances of the phlebotomists were further compared before and after the training program.

Results:

2622 from 2781 laboratories that did answer our survey used CLSI documents to standardize their procedures and process. The phlebotomists’ training for 20 days before our evaluation completely eliminated non-conformity procedures for: i) incorrect friction of the forearm, during the cleaning of the venipuncture site to ease vein location; ii) incorrect sequence of vacuum tubes collection; and iii) inadequate mixing of the blood in primary vacuum tubes containing anticoagulants or clot activators. Unfortunately the CLSI H03-A6 document does not caution against both unsuitable tourniquet application time (i.e., for more than one minute) and inappropriate request to clench the fist repeatedly. These inadequate procedures were observed for all phlebotomists.

Conclusion:

We showed that strict observance of the CLSI H03-A6 document can remarkably improve quality, although the various steps for collecting diagnostic blood specimens are not a gold standard, since they may still permit errors. Tourniquet application time and forearm clench should be verified by all quality laboratory managers in the services. Moreover, the procedure for collecting blood specimens should be revised to eliminate this source of laboratory variability and safeguard the quality.  相似文献   

14.

Introduction

We aimed to determine the number of repeated cholesterol (RC) tests and the ratio of unnecessary-repeated cholesterol (URC) tests among patients admitted to Pamukkale University Hospital (Denizli, Turkey) and provide solutions to avoid URC testing.

Materials and methods

Total cholesterol (T-cholesterol) tests (N = 86,817) between June 2014 and May 2015 were evaluated. The tests performed more than once per patient were determined as RC test (N = 28,811). RC test with an interval shorter than 4 weeks were determined as URC test (N = 3968) according to the shortest retest interval stated in ACC/AHA blood cholesterol guideline. RC testing included internal medicine, surgery and paediatric outpatients and inpatients. Reference change value (RCV) of total cholesterol was calculated.

Results

The 33.1% of the T-cholesterol tests were RC tests (N = 28,811), 13.7% of them were URC tests (N = 3968). Our RCV value was 25%. The percentage change between consecutive tests was less than RCV in 86.1% (N = 3418) of URC tests. URC tests were performed more frequently in patients with desirable total cholesterol value (P < 0.001).

Conclusion

There is a significant part of repeated T-cholesterol tests requested in our hospital. URC test requests can be evaluated by laboratories and the obtained data should be shared with clinicians. Laboratories can calculate RCV for the tests they performed and report this value with the test result. To prevent from URC tests, a warning plug-in can be added to hospital information software in accordance with guidelines to prevent from URC test requests.Key words: unnecessary repeated cholesterol test, cholesterol retest interval, reference change value, laboratory costs, preanalytical phase, test requesting  相似文献   

15.

Introduction

Total 25-hydroxyvitamin D [25(OH)D] is the most reliable indicator of vitamin D status. In this study, we compared two automated immunoassay methods, the Abbott Architect 25-OH Vitamin D assay and the Roche Cobas Vitamin D total assay, with the liquid chromatography-tandem mass spectrometry (LC-MS/MS).

Materials and methods

One hundred venous blood samples were randomly selected from routine vitamin D tests. Two of the serum aliquots were analyzed at the Abbott Architect i2000 and the Roche Cobas 6000’s module e601 in our laboratory within the same day. The other serum aliquots were analyzed at the LC-MS/MS in different laboratory. Passing-Bablok regression analysis and Bland-Altman plot were used to compare methods. Inter-rater agreement was analyzed using kappa (κ) analysis.

Results

The Roche assay showed acceptable agreement with the LC-MS/MS based on Passing-Bablok analysis (intercept: -5.23 nmol/L, 95% CI: -8.73 to 0.19; slope: 0.97, 95% CI: 0.77 to 1.15). The Abbott assay showed proportional (slope: 0.77, 95% CI: 0.67 to 0.85) and constant differences (intercept: 17.08 nmol/L; 95% CI: 12.98 to 21.39). A mean bias of 15.1% was observed for the Abbott and a mean bias of -14.1% was observed for the Roche based on the Bland-Altman plots. We found strong to nearly perfect agreement in vitamin D status between the immunoassays and LC-MS/MS. (κ: 0.83 for Abbott, κ: 0.93 for Roche) using kappa analysis.

Conclusion

Both immunoassays demonstrated acceptable performance, but the Roche Cobas assay demonstrated better performance than the Abbott Architect in the studied samples.Key words: 25-Hydroxyvitamin D, chromatography, immunoassay, methods, tandem mass spectrometry  相似文献   

16.

Introduction

To investigate the state of the art of intra-laboratory turnaround time (intra-TAT), provide suggestions and find out whether laboratories accredited by International Organization for Standardization (ISO) 15189 or College of American Pathologists (CAP) will show better performance on intra-TAT than non-accredited ones.

Materials and methods

479 Chinese clinical laboratories participating in the external quality assessment programs of chemistry, blood gas, and haematology tests organized by the National Centre for Clinical Laboratories in China were included in our study. General information and the median of intra-TAT of routine and stat tests in last one week were asked in the questionnaires.

Results

The response rate of clinical biochemistry, blood gas, and haematology testing were 36% (479 / 1307), 38% (228 / 598), and 36% (449 / 1250), respectively. More than 50% of laboratories indicated that they had set up intra-TAT median goals and almost 60% of laboratories declared they had monitored intra-TAT generally for every analyte they performed. Among all analytes we investigated, the intra-TAT of haematology analytes was shorter than biochemistry while the intra-TAT of blood gas analytes was the shortest. There were significant differences between median intra-TAT on different days of the week for routine tests. However, there were no significant differences in median intra-TAT reported by accredited laboratories and non-accredited laboratories.

Conclusions

Many laboratories in China are aware of intra-TAT control and are making effort to reach the target. There is still space for improvement. Accredited laboratories have better status on intra-TAT monitoring and target setting than the non-accredited, but there are no significant differences in median intra-TAT reported by them.Key words: quality indicators, quality control, clinical laboratory services  相似文献   

17.

Background

Due to the high prevalence of hemolysis in specimens received from the emergency department (ED), several strategies have been proposed to improve sample quality, but none of these seem effective to overcome the problem. In a preliminary study we showed that the use of S-Monovette blood collection system was effective to lower the risk of hemolysis in venous blood samples collected from intravenous catheters. This study was hence aimed to verify whether the replacement of a conventional vacuum system with S-Monovette may be effective to reduce the burden of hemolysis in the daily practice of a large urban ED.

Materials and methods

The study was divided in two observational periods of 4 months each. In the former period, blood was collected from intravenous catheters using BD Vacutainer SST II Plus plastic serum tubes, whereas in the latter period the blood was drawn from intravenous catheters using S-Monovette blood tubes in aspiration mode. Sample hemolysis was automatically assessed in all serum samples by photometrical measurement.

Results

The total number of hemolysed serum specimens was 624/14155 (4.41%) in the first phase of the study, and 342/13319 (2.57%) in the second phase of the study (P < 0.001).

Conclusion

Results of our study confirm that the introduction of the Sarstedt S-Monovette blood tubes has reduced the hemolysis rate in the emergency department compared to the previously used BD Vacutainer® SST II Plus plastic serum tubes.Key words: preanalytical phase, hemolysis, blood specimen collection, catheter, emergency department  相似文献   

18.

Introduction

The study was performed to compare and analyze the inter-departmental variability in the request of rarely requested laboratory tests in primary care, as opposed to other more common and highly requested tests.

Materials and methods

Data from production statistics for the year 2012 from 76 Spanish laboratories was used. The number of antinuclear antibodies, antistreptolysin O, creatinine, cyclic citrullinated peptide antibodies, deaminated peptide gliadine IgA antibodies, glucose, protein electrophoresis, rheumatoid factor, transglutaminase IgA antibodies, urinalysis and uric acid tests requested was collected. The number of test requests per 1000 inhabitants was calculated. In order to explore the variability the coefficient of quartile dispersion was calculated.

Results

The smallest variation was seen for creatinine, glucose, uric acid and urinalysis; the most requested tests. The tests that were least requested showed the greatest variability.

Conclusion

Our study shows through a very simplified approach, in a population close to twenty million inhabitants, how in primary care, the variability in the request of laboratory tests is inversely proportional to the request rate.Key words: primary care, laboratory proficiency testing, clinical laboratory services, test requesting, preanalytical phase  相似文献   

19.

Introduction:

We hypothesized that patients are poorly informed about proper procedure for 24-hour urine specimen collection and its relevance in determination of biochemical analytes, despite availability of leaflets and webpage with instruction for collection. The aim of this survey was to question outpatients how well are they informed about procedure of 24-hour urine specimen collection.

Materials and methods:

The survey with 10 questions was done in outpatient laboratory of University Department of Chemistry, Medical School University Hospital Sestre Milosrdnice, Zagreb, Croatia. The study included 59 patients with collected 24-hour urine sample who have consented to participate in the survey.

Results:

Out of 59 participants, most of them (0.97) were older than 40 years. Internet was not recognized as a source of information (1/59). Almost one third of the patients have changed their drinking habits to collect more urine volume. Although most of the patients (0.60) were aware that the bottle of water is the best choice for the container, almost half of them were collected urine samples in the plastic soft drink bottle. Laboratory staff and physicians often have given information about proper collection procedure, but that information was insufficient.

Conclusions:

Patients are usually not aware of importance of proper preanalytical procedure for collecting urine specimen and how improper collection could affect results of requested tests. Education of outpatients, general practitioners and laboratory staff is needed in order to improve sample quality and trueness of results.  相似文献   

20.

Introduction:

Sample type recommended by the manufacturer for the digoxin Abbott assay is either serum collected in glass tubes or plasma (sodium heparin, lithium heparin, citrate, EDTA or oxalate as anticoagulant) collected in plastic tubes. In our hospital samples are collected in plastic tubes. Our hypothesis was that the serum sample collected in plastic serum tube can be used interchangeably with plasma sample for measurement of digoxin concentration. Our aim was verification of plastic serum tubes for determination of digoxin concentration.

Materials and methods:

Concentration of digoxin was determined simultaneously in 26 venous blood plasma (plastic Vacuette, LH Lithium heparin) and serum (plastic Vacuette, Z Serum Clot activator; both Greiner Bio-One GmbH, Kremsmünster, Austria) samples, on Abbott AxSYM analyzer using the original Abbott Digoxin III assay (Abbott, Wiesbaden, Germany). Tube comparability was assessed using the Passing Bablok regression and Bland-Altman plot.

Results:

Serum and plasma digoxin concentrations are comparable. Passing Bablok intercept (0.08 [95% CI = −0.10 to 0.20]) and slope (0.99 [95% CI = 0.92 to 1.11]) showed there is no constant or proportional error.

Conclusion:

Blood samples drawn in plastic serum tubes and plastic plasma tubes can be interchangeably used for determination of digoxin concentration.  相似文献   