BOD5稀释水曝气方法对比及鼓气曝气法参数研究
引用本文:杨保华,陈剑虹.BOD5稀释水曝气方法对比及鼓气曝气法参数研究[J].实验室研究与探索,2004,23(6):37-38.
作者姓名:杨保华  陈剑虹
作者单位:长沙环境保护职业技术学院,湖南,长沙,410004
摘    要:对鼓气曝气和抽气曝气两类曝气方法做了对比研究,结果显示:抽气曝气,稀释水DO始终达不到8mg/L,开放静置24h,DO几乎不增加;鼓气曝气30min,稀释水DO已稳定超过8mg/L,鼓气曝气120min后开放静置20h,DO仍有8mg/L。本文认为,一般情况下,BOD5稀释水曝气只能采用鼓气曝气法;在排量≥100L/min,水量≤20L条件下,曝气120min后,24h内,稀释水DO可满足技术要求。

关 键 词:BOD5  稀释水  鼓气曝气
文章编号:1006-7167(2004)06-0037-02
修稿时间:2003年3月19日

Comparison of Aeration Methods for Diluted Water and Study on Parameters of Blower-Aeration
YANG Bao-hua,CHEN Jian-hong.Comparison of Aeration Methods for Diluted Water and Study on Parameters of Blower-Aeration[J].Laboratory Research and Exploration,2004,23(6):37-38.
Authors:YANG Bao-hua  CHEN Jian-hong
Abstract:Two methods of blower-aeration and vacuum pump-aeratin were comparatively studied. With vacuum pump-aeratin, DO of diluted water can not approach to 8mg/L all the time and increases hardly in 24h open and quiet next .With blower-aeration, DO of diluted water exceedes 8mg/L after 30 min and keep 8mg/L over in 24h open and quiet next after 120 min. Blower-aeration is the only method for aeration of diluted water. 120min is sufficient for blower-aeration and diluted water can be used in 24h.
Keywords:BOD_5  diluted water  blower-aeration  